APPENDIX E - CE CONFORMANCE

FLUID COMPONENTS INTL

MT91

All options conform to CE marking
requirements.

X Y

X or Y options only conform to CE marking

All other configurations must be approved for CE marking conformity by FCI's Engineering Department.

1-15

2

16

17-18

Appendix E. CE Conformance

Approved CE Marking For The MT91 Configuration

Approved Options

1) all insertion U-lengths

2) all specified application combinations

3) 85 to 260 VAC power input.

4) all metallic sensor elements

5) all cable jackets and lengths

6) Aluminum and 300 series stainless steel NEMA 4X local and remote applications.

7) All process connections and flanges

8) All process temperature ranges

Approved Part Numbers

Conditionally Approved CE Marking

24 VDC Factory Selectable Options

Input power of 24 VDC may be optionally selected. Since the manufacturer does not supply the power source or
transformer for these connections, the responsibility for the conditioning of these sources and associated compliance
to the EMC Directive shall be the responsibility of the user.

Panel and Rack Mounted Control Circuit Options

Conditionally approved models included in this family are supplied with panel or rack mounted electronic assemblies
without a remote enclosure (MT91 block 16, codes 1 or 3). Since the manufacturer does not supply an EMC
enclosure for these configurations the associated compliance to the EMC directive shall be the responsibility of the
user.

Installation Conformity Criteria

Grounding

The enclosure must be grounded to earth ground through a path of less than 1 ohm.

Interconnecting Cables

All interconnecting cables between the local enclosure, power source and monitoring device shall be enclosed in
metal conduit. AC power input cabling shall be enclosed separately in conduit before entrance into the instrument
and is not to be combined with switch or monitor output cabling.