Model MT91 Multipoint Flowmeter

CHAPTER 5 - TROUBLESHOOTING

FLUID COMPONENTS INTL

5. Troubleshooting

Warning:

Only qualified personnel should attempt to test this device. The operator assumes all
responsibilities for safe practices while troubleshooting.

Caution:

The flow transmitter contains electrostatic discharge (ESD) sensitive devices. Use standard ESD
precautions when handling the flow transmitter. See Chapter 2, Installation for ESD details.

Quick Check

Use Table 5-1 as a quick check of problems and solutions. More in-depth discussions follow this table.

Table 5-1. Quick Check Troubleshooting

Problem

Solution

No Display

Check fuses. Verify correct power is applied.
Verify the ribbon cable between the backplane board J18 and the display J1 is
solidly connected and the red stripe is at pin 1. See Figures 2-3 and 2-4.

No Display or Dim Display

Adjust potentiometer R1 (located on display board) clockwise (10 turns
maximum) until display appears. See Figures 2-3.

Display is locked up

Cycle the input power or press any key for more than 10 seconds.

Incorrect Readings

Verify that all ribbon cables are firmly seated. See Figure 2-3 and 2-4.

Non-maintenance Observations

At this point, simply observe the system setup to verify operation. No disassembly or testing is required at this time.

Check Serial Numbers

Verify that the serial number of the flow elements and the flow transmitter are the same. The flow elements and the
flow transmitter are a matched set and cannot be operated independently of each other.

Check Input Power

Verify that the correct power source is turned on and connected.

Check the Unit Installation

Review the information on instrument installation in Chapter 2 to verify correct mechanical and electrical
installation. Verify that all ribbon cables are seated firmly.

Relay Output Board J1

Þ

Backplane Board J21 and/or J22

Analog Output Board J1

Þ

Backplane Board J19

Display Board J1

Þ

Backplane Board J18

Check for Moisture

Check for moisture on the flow transmitter. Moisture on the flow transmitter may cause intermittent operation.