ST98 HART Option

Installation and Operation Guide

This manual is used to augment other instrument manuals when the HART option is used. NOTE: ST98 HART supports “Normal
Mode” and it does not support “Burst Mode”.

Step 1.

Installation

Use the following steps to install the HART option for the ST98:

1. All the FCI transmitters that incorporate the HART protocol are set at the factory to polling address “0”. Users should reset the

polling address to match the needs of their network.

2. At polling address “0”, the ST98 outputs a current of 4 - 20mA; at any other polling address, the ST98 outputs a fixed 4mA

current; in compliance with the HART standard. A digital signal is provided with all polling addresses.

3. If the ST98 is going to be used as a HART network transmitter, the polling address needs to be changed to other than “0”.

4. All the FCI transmitters that incorporate the HART protocol are configured by the factory as a current source; and JP1 through JP3

are set for “Internal Power” by the factory.

5. When the ST98 is connected in a HART Network a recommendation is the instrument be configured for an external current source

by resetting JP1, JP2, and JP3 on the Input/Output Board, and the polling address must be changed to other than “0”. The power
supply feeding the network must provide sufficient current to support the instruments in the network.

Output Current Jumper Settings

Ext

JP1

JP2

JP3

Ext

JP1

JP2

JP3

External Power

Internal Power

Jumpers

ST98 Input/Output Board

Internal Power setting shown

Notice of Proprietary Rights

This document contains confidential technical data, including trade secrets and proprietary information which is the property of Fluid Components
International LLC (FCI). Disclosure of this data to you is expressly conditioned upon your assent that its use is limited to use within your company only
(and does not include manufacture or processing uses). Any other use is strictly prohibited without the prior written consent of FCI.

Visit FCI on the Worldwide Web: www.fluidcomponents.com

1755 La Costa Meadows Drive, San Marcos, California 92078 USA - 760-744-6950 - 800-854-1993 - Fax 760-736-6250

Advertising