2551 Magmeter
Pipe Size
(Metric)
Fitting
Type
K-Factor
Gallons
K-Factor
Liters
20 mA=
in GPM
20 mA=
in LPM
POLYPROPYLENE FITTINGS (DIN/ISO, BS, ANSI)
DN15
PPMT005
2192.73
579.32
13.68
51.78
DN20
PPMT007
1327.81
350.81
22.59
85.52
DN25
PPMT010
737.16
194.76
40.70
154.04
DN32
PPMT012
453.46
119.81
66.16
250.41
DN40
PPMT015
275.03
72.66
109.08
412.86
DN50
PPMT020
164.17
43.35
182.74
691.66
PVDF FITTINGS (DIN/ISO, BS, ANSI)
DN15
SFMT005
1946.49
514.26
15.41
58.34
DN20
SFMT007
1158.05
305.96
25.91
98.05
DN25
SFMT010
749.09
197.91
40.05
151.58
DN32
SFMT012
439.51
116.12
68.26
258.36
DN40
SFMT015
248.93
65.77
120.52
456.16
DN50
SFMT020
146.85
38.80
204.30
773.26
PVC FITTINGS (DIN/ISO, BS, ANSI)
DN15
PVMT005
2067.76
546.30
14.51
54.91
DN20
PVMT007
1136.61
300.29
26.39
99.90
DN25
PVMT010
716.52
189.31
41.87
158.47
DN32
PVMT012
446.07
117.85
67.25
254.56
DN40
PVMT015
278.83
73.67
107.59
407.23
DN50
PVMT020
159.36
42.10
188.26
712.55
Plastic Installation Fittings for Metric Pipes:
Polypropylene True Union Tees and Wafers
PVDF True Union Tees
PVC True Union Tees
10. Calibration Data: K-factors and Full Scale Current Values