Signet 2724-2726 and 2734-2736 Series pH and ORP Electrodes

The Signet pH/ORP electrodes are designed to install in tanks by attaching conduit to the ¾ in. threads at the top of the
accompanying preamplifi er or sensor electronics:
1. The O-ring at the top of the electrode fi ts very tightly into the preamplifi er. Use a small amount of non-petroleum based

lubricant to assist the assembly.

2. To prevent moisture from migrating into the preamplifi er, backfi ll the conduit with 3 in. to 4 in. of sealant.
3. Mount electrodes in a location with ample clearance to remove them for periodic cleaning and recalibration.
4. Choose a location that keeps the electrode glass completely submersed at all times.

Technical Note:
• Mount the electrode near tank outlets,

away from reagent addition areas.

• Use the ¾ in. threads at the top of the preamplifi er

to run the cable inside pipe or conduit.

• Place the electrode tip in pH 4 buffer during system

maintenance or storage to avoid dehydration.

Type

Description

Type

Description

Plastic tees

PVC saddles

Iron
strap-on
saddles

‡$YDLODEOHLQòLQWRLQVL]HV
‡39&&39&ZVROYHQWFHPHQWVRFNHW
‡39')RU33ZXQLRQHQGILWWLQJV

‡$YDLODEOHLQLQWRLQVL]HV
‡5HTXLUHVLQKROHLQSLSH

‡$YDLODEOHLQLQWRLQVL]HV
‡5HTXLUHVLQKROHLQSLSH

‡$YDLODEOHLQLQWRLQVL]HV
‡5HTXLUHVLQKROHLQSLSH
‡,QVWDOOE\FHUWLILHGZHOGHURQO\

‡$YDLODEOHLQòLQWRLQVL]HV
‡)HPDOH137HQGV

‡8VHIRULQVWDOODWLRQLQSLSHV!LQ óLQ137
‡39&&39&RU39')YHUVLRQV
‡6SHFLI\VRFNHWRUóLQFK137PDOHWKUHDGV

&DUERQVWHHO
ZHOGROHWV

&DUERQVWHHO
WKUHDGHGWHHV

8QLYHUVDO
pipe adapters

Signet Installation Fittings

Submersible Installations

Lubricate
O-rings

Fill with
3 to 4 in.
of sealant

1

2

Open
tanks

Closed

tanks

3

3

4

Removing from In-Line Installations

In-line removal Instructions:

1. Depressurize and vent the piping system.
2. Drain the system to below sensor level.
3. Wear safety goggles or face shield during removal.

Use all appropriate eye, face, hand, body and/or respiratory
protection when working with chemicals or solvents.

4. Place a Lockout tag on the pipe when the sensor is

removed for maintenance to prevent accidental opening
and exposure to potentially hazardous chemicals.

• The use of this product assumes

that operators are trained and are
familiar with this type of device.

• They should be knowledgeable of

the potential risks associated with
pressurized piping systems.

• Operators MUST follow all

necessary safety procedures.

The user must supply the following hardware
to complete a submersible installation:

¾ in. NPT threaded pipe or conduit

Wiring junction box

Pipe clamps (quick-release type recommended)

• Tank

fl ange for closed tanks

Submersible Installation

2724-2726 / 2734-2736 with 2750/2760 Preamplifi er