Channel selection keys, Arrow keys, On key – Hach-Lange BODTRACK II User Manual User Manual

Page 11: Off key, Startup, Turn the instrument on, Set the clock, Channel selection keys arrow keys on key off key, Turn the instrument on set the clock

Advertising
background image

Channel selection keys

Push the related channel selection key to show data for one of the six
bottles. The channel selection keys are also used in the instrument setup
menu to select a parameter to be edited. Refer to

Figure 2

on page 8

and

Table 3

.

Table 3 Channel key setup parameters

Channel

Parameter

1

Year (0–99)

2

Month (1–12)

3

Day (1–31)

4

Hour (0–24)

5

Minute (0–59)

6

Test length (5, 7 or 10 days)

Arrow keys

The display shows a graph of BOD values on the vertical axis and time
in days on the horizontal axis. Push the LEFT and RIGHT arrows to
move the cursor along the BOD curve to show the approximate
coordinates (time, BOD) of the selected data point.
The time interval and BOD value of the data point are shown in the lower
right of the display. The cursor is automatically placed at the most
recently collected data point in a channel display.
Push and hold the LEFT and RIGHTarrows at the same time to go into
the instrument setup menu. The arrow keys are also used to change the
time, date, test length and range.

ON key

To go to the range selection menu, push ON from a channel display
screen. Then push and hold ON to start the test for the selected channel.

OFF key

When a test is in DELAY or RUN modes, push and hold OFF to
manually end the test. The instrument will show END. The OFF key is
also used to exit the instrument setup menu or range selection menu.
The changes made before the menu is exited are saved.

Startup

Turn the instrument on

Note: The ON and OFF keys are used to start and stop a test. They do not power
the instrument on and off.

1. Connect the power adaptor to the DC power connector (

Figure 1

on page 8).

2. Toggle the power switch to set the instrument to on and off. (

Figure 1

on page 8).

Set the clock

All the channels must show END or CLEAR before the clock can be set.

1. Push and hold the two arrow keys at the same time until the

instrument setup menu is shown.

2. Push the applicable channel key to select the clock parameter to be

adjusted.

3. Use the arrow keys to edit the selected parameter. Adjust each

parameter in the same manner.

4. When all the time adjustments are complete, push OFF to save and

go back to the data display screen.

English 9

Advertising