bottle fence. Move the tubing counter-clockwise and secure the bottle
cap inside the fence.

Troubleshooting

Incorrect BOD curves

Figure 4

shows incorrect BOD curves for a 10 day test period. For typical

curves, refer to

Typical curves

on page 17.

Figure 4 Incorrect curves

1 High oxygen demand

4 Initial sample temperature below

20 ºC or supersaturated with
oxygen

2 Nitrification

5 Bottle leak

3 Excessive time lag

High oxygen demand

Refer to

Figure 4

on page 20. Samples that are above range (for

example, a BOD over 350 mg/L when a 160-mL sample is taken) will

20 English