Hach-Lange BODTRACK II User Manual User Manual

Page 18

Advertising
background image

7. Push and hold the
LEFT and RIGHT
arrows at the same time
to go to the instrument
setup menu.
Note: Set the time and
date if necessary (

Set

the clock

on page 9).

8. Push the channel
6 key to go to the test
length parameter. Use
the arrow keys to select
a 5, 7 or 10 day test.
Note: The selected test
length is for all six
channels.

9. Push OFF to save
selections and exit the
menu.

10. To start the test,
push the channel
number for the bottle.

11. Push ON. The
range selection menu is
shown.

12. Use the arrows to
select the test range.
Note: Use the left arrow
key for the 0 to 35 and
0 to 70 mg/L ranges.
Use the right arrow key
for the 0 to 350 and 0 to
700 mg/L ranges.

13. Push and hold ON
to start a test. A graph
will show.
Note: To cancel a test,
push and hold OFF.
There is a built-in
1 hour
instrument/sample
equilibration period
before data collection.
The display shows
"DELAY" during this
period.

14. Do steps
10 through 13 again to
set the test range and
start each of the six
channels. It is not
necessary to operate all
six channels if less than
six samples are
available.

16 English