Clean the instrument, Clean the sample bottles, Clean the stir bars and seal cups – Hach-Lange BODTrak II Basic User Manual User Manual

Page 10: Storage, Troubleshooting, Incorrect bod curves

Advertising
background image

C A U T I O N

Chemical exposure hazard. Obey laboratory safety procedures and
wear all of the personal protective equipment appropriate to the
chemicals that are handled. Refer to the current material safety data
sheets (MSDS) for safety protocols.

Clean the instrument

Clean spills on the instrument with a soft cloth that has been dampened
with deionized or distilled water.

Clean the sample bottles

Clean the sample bottles and caps with a brush, water and a mild
detergent. Flush the containers with fresh water followed by a distilled
water rinse.

Clean the stir bars and seal cups

Clean the stir bars with hot water and soap. Use a brush to remove
deposits. Rinse with fresh water and then rinse with distilled water.
Carefully empty and rinse the seal cups with water. Invert to dry.

Storage

The bottle fences prevent tipping of the bottles and provide tubing
management in storage. For storage, put the tubing in the opening in the
bottle fence. Move the tubing counter-clockwise and secure the bottle
cap inside the fence.

Troubleshooting

Incorrect BOD curves

Figure 4

shows incorrect BOD curves for a 10 day test period. For typical

curves, refer to

Typical curves

on page 7.

Figure 4 Incorrect curves

1 High oxygen demand

4 Initial sample temperature below

20 ºC or supersaturated with
oxygen

2 Nitrification

5 Bottle leak

3 Excessive time lag