Messages.message in.show" item, Messages.messages out.state" item, Messages.message out.errors" item – Hach-Lange ORBISPHERE 6110 OPC Instruction Manual User Manual
Page 13

Note: The package must be created directly on the 6110 instrument, or through a web browser connected to the
6110 web server. It cannot be created using the OPC link
"Messages.Message In.Show" item
This item shows the specified message on the instrument.
"Messages.Messages Out.State" item
This output item indicates when the package can be changed for a new measurement analysis.
State value
Stet description
Ready
The package ID can be set in order to start a
measurement analysis.
Note: Check the "Message Out.Errors" item to check if the
analysis is able to start (see
on page 11).
Busy
The system is busy (e.g. measurement in progress,
configuration being edited, etc.), preventing a
measurement analysis being started.
"Messages.Message Out.Errors" item
This output item provides information about any error(s) associated with the system. An error
prevents the instrument from running a measurement analysis. The instrument requires attention to
correct the fault. This is a flag field where the indicated value is computed by the addition of all active
errors.
Error value
Error description
0x00000000
No error
0x00000001
EC board
0x00000002
ADIO board
0x00000004
CTRL board
0x00000008
O2 sensor
0x00000010
O2 sensor residual
0x00000020
O2 sensor noisy
0x00000040
TC chip
0x00000080
Temperature sensor
0x00000100
Flow sensor
0x00000200
Pressure sensor
0x00000400
Foam detector
0x00000800
Electrovalves
0x00001000
Motor
0x00002000
Ultrasound
0x00004000
Protection cover
0x00008000
Front door
0x00010000
Emergency stop
0x00020000
Sensor status
English 11