Error value

Error description

0x00040000

Antifoam counter

0x00080000

Pending action

0x00100000

Pending manual action

0x00200000

No flow

0x00400000

RIM detector error

0x00800000

Upper limit

0x01000000

System initialization required

0x02000000

Communication error

0x04000000

ADIO board KO

0x08000000

CTRL board KO

0x10000000

EC board KO

0x20000000

CO2 foam out of specifications

0x40000000

Foam too many times

"Messages.Message Out.Warnings" item

This output item provides information about any warning(s) associated with the system. A warning
does not prevent the instrument from running a measurement analysis. This is a flag field where the
indicated value is computed by the addition of all active warnings.
The list items are identical to those listed in the corresponding errors table (see

"Messages.Message

Out.Errors" item

on page 11).

"Messages.Message Out.Message In Last Error" item

This item indicates the last error raised by an incoming OPC message. For example, when a
"Message In.New Package" item (see

"Messages.Message In.New Package" item

on page 10)

requests setting the package to one that is not defined on the 6110 instrument, the error value will be
"POOR".

Note: In the following table "par_name" represents the parameter name to which the error relates.

Error value

Error description

NONE

No error

UNKN par_name

Unknown parameter

POOR par_name

Parameter out of range

TYPE par_name

Parameter wrong type

DENY par_name

Parameter access denied

NR par_name

Not ready, a command has been received at the wrong
time