16

Operation

5.5 The commands under SENSOR SETUP

SENSOR SETUP

SELECT SENSOR (for several sensors)

CALIBRATE

1 POINT BUFFER

Calibration using a buffer — normally pH 7, see

5.6.1 Calibration using

a buffer solution

.

2 POINT BUFFER

Calibration using two buffers — normally pH 7 and pH 4 or 10, see

5.6.2 Calibration using two buffer solutions

.

1 POINT SAMPLE

Calibration using a known sample, see

5.6.3 Calibration using a

sample solution

.

2 POINT SAMPLE

Calibration using two samples, both with known pH value, see

5.6.4

Calibration using two sample solutions

.

PROCESS TEMP

Set temperature displayed.

CAL CONFIG

OUTPUT MODE

Choose between:
ACTIVE, measured value and thus the outputs signals follow
the calibration process;
HOLD, the last measured value and thus the output signal
are frozen;
TRANSFER, the measured value and thus the output signal are set
to a defined value and
CHOICE, after each calibration you can choose one of the three above
options.

CAL REMINDER

You can set when the next calibration is to be performed. The
controller then automatically indicates when the next calibration is due.

DEFAULT SETUP

Return to the factory settings after prompt for confirmation.

CONFIGURE

EDIT NAME

Enter a 10-character name and accept.

SELECT MEASURE

Choose between pH and REDOX.

TEMP UNITS

Temperature figures in degrees Celsius or Fahrenheit.

LOG SETUP

LOG INTERVAL

Choose between the values available or DISABLED.

TEMP. INTERVAL

Choose between the values available or DISABLED.

FILTER

Enter a value.

DISPLAY FORMAT

Choose the display format.

SELECT BUFFER

Choose a buffer.

T-SENSOR

AUTOMATIC
MANUAL

Enter a value.

MAINS FREQ.

Enter the mains frequency.

DEFAULT SETUP

Returns to the factory settings after a prompt for confirmation.

DIAG/TEST

PROBE INFO

Provides information on driver, software and serial number

CAL DATA

Provides information on the offset (T), slope und offset (pH).

SIGNALS

Provides information on the raw measured data in mV, pH and °C / F

COUNTERS

Counts up after each calibration and compares the value with the
setting made under CAL REMINDER

TEST/MAINT

Disable OUTPUT during test or maintenance

SERVICE

Reset Service Counter