22
Maintenance
The new electrode is fitted in the reverse order of removal.
1. Slide the new electrode into the sensor until the base ring is in contact with
the electrode so that the sensor is sealed.
2. Calibrate the sensor using two buffer solutions (see
5.6.2 Calibration using
two buffer solutions
).