Section 7 faults, causes, rectification, 1 error messages, 2 warnings – Hach-Lange 1200-S sc User Manual

Page 25: 1 error messages 7.2 warnings, Occurred (see section

Advertising
background image

23

Section 7

Faults, causes, rectification

7.1 Error messages

Possible sensor errors are displayed by the controller.

7.2 Warnings

Possible warning messages are displayed by the controller.

Table 1: Error messages

Error displayed

Cause

Rectification

*****

No communication with the
controller

Check the connection to the controller
Check the cable to the controller
Check the 12 V power supply

SENSOR MISSING
FFFFFFFFFFFFF

No communication with the
controller

Check the connection to the controller
Check the cable to the controller
Check the 12 V power supply

TEMP TOO LOW

Measured temperature < –5 °C

Ensure that the medium temperature is > –5 °C.
Check whether the internal resistance of the Pt 100 is
approx. 99 Ohm.

TEMP TOO HIGH

Measured temperature > +100 °C

Ensure that the medium temperature is < +100 °C.
Check whether the internal resistance of the Pt 100 is
approx. 138.5 Ohm.

pH TOO LOW

pH is lower than –2 pH

Check the sensor for leaks.
Change the electrode.
Change the board.

pH TOO HIGH

pH is higher than +14 pH

Check the sensor for leaks.
Change the electrode.
Change the board.

mV TOO LOW

mV is lower than –1500 mV

Check the sensor for leaks.
Change the electrode.
Change the board.

mV TOO HIGH

mV is higher than +1500 mV

Check the sensor for leaks.
Change the electrode.
Change the board.

Table 2: Warnings

Error displayed

Cause

Rectification

CAL TOO OLD

The last calibration was more than x days ago.
(Setting from the sensor setup)

Calibrate the sensor.
Set the calibration interval in the sensor setup.

HUMIDITY BAG

The desiccant bag is more than 1000 days old.

Please contact service.

REF IMP. LOW

Impedance of Reference electrode too low

Please contact service.

REF IMP. HGH

Impedance of Reference electrode too high

Please contact service.

GLASS IMP. LOW

Impedance of Glass electrode too low

Please contact service.

GLASS IMP. HGH

Impedance of Glass electrode too high

Please contact service.