User manual, Specifications, English – Hach-Lange POLYMETRON 83xx Basic User Manual User Manual

Page 3

Advertising
background image

Specifications

Specifications are subject to change without notice.

Table 1 Probe specifications

8310 / 8315

8311 / 8316

Applications

Pure and ultra pure water

Moderately conductive solutions

K (cm

-1

)

0.01

0.1

Accuracy

< 2%

< 2%

Transmitter measurement range

0.01 to 200 μS.cm

-1

0.1 μS to 2 mS.cm

-1

Pt100 temperature response (t
90%)

< 30 seconds

< 45 seconds

Maximum temperature (°C)

125 (8310)
150 (8315)

125 (8311)
150 (8316)

Maximum pressure (bars)

10 (8310)
25 (8315)

10 (8311)
25 (8316)

Sample connection

¾ inch NPT

¾ inch NPT

8312 / 8317

8394

Applications

Waste water and drinking water

Food and pharmaceutical industries

(sterilized)

K (cm

-1

)

1.0

0.01

Accuracy

< 2%

< 2%

Transmitter measurement range

1 μS to 20 mS.cm

-1

0.01 to 200 μS.cm

-1

Pt100 temperature response (t
90%)

< 3 minutes

< 45 seconds

Maximum temperature (°C)

125 (8312)
150 (8317)

150

Maximum pressure (bars)

10 (8312)
25 (8317)

25

Sample connection

¾ inch NPT

Tri-Clamp 1½ or 2 inch

Table 2 Flow chamber specifications

08313=A=0001

08318=A=0001

Material

PVC

316 L SS

Maximum temperature (°C)

60 at 2 bar

150

Maximum pressure (bars)

15 at 25°C

25

Sensor connection

¾ inch NPT

¾ inch NPT

Sample connection

¾ inch NPT

¼ inch NPT

English 3