Note: Refer to the user manual delivered with the transmitter for a detailed description of the cable connectors on
the transmitter.

Probe installation

In

Figure 5

on page 7,

Figure 6

on page 7 and

Figure 7

on page 8 the annotations A, B and

C indicate:

A: Ideal installation - perfect immersion of the electrode surfaces.
B: Good installation - satisfactory immersion of the electrode surfaces.
C: Poor installation - incomplete immersion of the electrodes, the conductivity will be too low.

On piping

Immerse the internal electrode completely in the process sample. For a 90° installation, take into
account the dimensions (see

Dimensions

on page 5).

Note: In the following illustrations, the arrows indicate the sample flow direction.

Installation example for the 8315 probe

Figure 5 8315 Probe

Installation example for the 8394 probe
This probe installs perfectly in a Tri-Clover

®

Tri-Clamp

Tee starting from 1.5 inch diameter (A), and

also at a 90° angle starting from 2 inches (B). All Tri-Clamp

Tees are in compliance with 3A

standards for Cleaning In Place (CIP).

Figure 6 8394 Probe