Note: Refer to the user manual delivered with the transmitter for a detailed description of the cable connectors on
the transmitter.
Probe installation
In
on page 7,
on page 7 and
on page 8 the annotations A, B and
C indicate:
• A: Ideal installation - perfect immersion of the electrode surfaces.
• B: Good installation - satisfactory immersion of the electrode surfaces.
• C: Poor installation - incomplete immersion of the electrodes, the conductivity will be too low.
On piping
Immerse the internal electrode completely in the process sample. For a 90° installation, take into
account the dimensions (see
on page 5).
Note: In the following illustrations, the arrows indicate the sample flow direction.
Installation example for the 8315 probe
Figure 5 8315 Probe
Installation example for the 8394 probe
This probe installs perfectly in a Tri-Clover
®
Tri-Clamp
™
Tee starting from 1.5 inch diameter (A), and
also at a 90° angle starting from 2 inches (B). All Tri-Clamp
™
Tees are in compliance with 3A
standards for Cleaning In Place (CIP).
Figure 6 8394 Probe