Cable connections – Hach-Lange POLYMETRON 83xx Basic User Manual User Manual

Page 6

Advertising
background image

Model

h max (mm)

H min (mm)

D min (standard piping)

8316

28

80

DN50 or 2 inch

8317

28

90

DN75 or 3 inch

8394

21.5

65.5

DN50 or 2 inch

Figure 3 Flow chamber dimensions

Cable connections

C A U T I O N

Connect the cable quickly to avoid any risk of humidifying the connector.

Figure 4 Cable connections

Note: For illustration purposes, cable part number 08319=A=00xx is shown in

Figure 4

.

1 External shielding

4 External electrode

2 Internal shielding

5 Pt100

3 Internal electrode

6 Pt100

The cable is available in lengths of 5, 10 or 20 meters and must be connected in compliance with the
following table:

Function

Color

External shielding

White (red tip)

Internal shielding

White (orange tip)

Internal electrode

White (yellow tip)

External electrode

Red

Pt100

Black

Pt100

Blue

6 English