Reset calibration options, Maintenance, Cleaning and decontamination – Hach-Lange POLYMETRON 9586 User Manual

Page 31: Fuse replacement, Battery replacement, Troubleshooting, General troubleshooting

Advertising
background image

3. Select CAL OPTIONS and push enter.
4. Use the arrow keys to select an option and push enter.

Option

Description

CAL REMINDER Sets a reminder for the next calibration in days, months or years—select the required

delay from the list.

OP ID ON CAL

Includes an operator ID with calibration data—YES or NO (default). The ID is entered
during the calibration.

Reset calibration options

This option is reserved for Hach Lange service technicians.

Maintenance

D A N G E R

Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.

Cleaning and decontamination

D A N G E R

Always remove power from the controller before performing maintenance activities.

The analyzer does not normally require any cleaning or decontamination. If needed, clean the
exterior of the instrument with a moist cloth and a mild soap solution. Never use cleaning agents
such as turpentine, acetone or similar products to clean the instrument, including the display and any
accessories.

Fuse replacement

Fuses are not user-serviceable items. The need for fuse replacement in controllers indicates severe
technical failure and is therefore considered to be a service activity. If a blown fuse is suspected,
contact Technical Support.

Battery replacement

The lithium ion backup battery is not user replaceable. Contact Technical Support for replacement.

Troubleshooting

General troubleshooting

Table 6 Controller

Problem

Resolution

No current output

Verify current output configuration.

Test current output signal using the Test/Maintenance submenu.
Input a current value and verify the output signal at the controller
connections.

Contact Technical Support.

Incorrect current output

Verify current output configuration.

Test current output signal using the Test/Maintenance submenu.
Input a current value and verify the output signal at the controller
connections. If the output is incorrect, perform an output
calibration.

English 29