Table 8 Sensor DIAG/TEST menu

Option

Description

MODULE INFORMATION

Shows information about the sensor module.

SENSOR INFORMATION

Shows the name and serial number that was entered
by the user.

CAL DAYS

Shows the number of days since the last calibration.

CAL HISTORY

Shows a list of all calibrations by date/time stamp. Use
the arrows keys to select a calibration and push enter
to view the details.

RESET CAL HISTORY

Resets the calibration history for the sensor (requires
service-level passcode). All previous calibration data is
lost.

POLARIZATION

Contacting conductivity sensors only. Shows
information about the electrode polarization, the cable
capacitance and the time before the next
measurement.

SENSOR SIGNALS

Shows the current sensor signal information.

MEMBRANE DAYS

Shows the number of days that the membrane has
been in operation.

RESET MEMBRANE

Resets the number of days that the membrane has
been in operation and resets all calibration data to
defaults.

Warning and error conditions

Follow the steps below to acknowledge controller warnings.

1. Push the menu key and select DIAGNOSTICS.
2. Select the device (controller, sensor, network card) with the warning or error and push enter.
3. Select the warning, error or event list and push enter.
4. Select YES and push enter to acknowledge the warning.

Note: Errors cannot be acknowledged.

Warning list

A warning icon consists of an exclamation point within a triangle. Warning icons appear on the right
of the main display below the measurement value. A warning does not affect the operation of menus,
relays and outputs. To view warnings, push the menu key and select DIAGNOSTICS. Then select
the device to view any problems associated with that device. The warning icon will no longer be
displayed once the problem has been corrected or acknowledged.
A list of possible warnings is shown in

Table 9

.

Table 9 Warning list for oxygen scavenger sensors

Warning

Description

Resolution

O2 SCAVENGER TOO HIGH

The measured value is >
10000 ppm

Make sure that the oxygen
scavenger (hydrazine or
carbohydrazide) level in the process
water is within the operating limits of
the sensor. Calibrate or replace the
sensor.

O2 SCAVENGER TOO LOW

The measured value is < 0 ppm

Calibrate or replace the sensor.

TEMP TOO HIGH

The measured temperature is >
50 °C

Reduce sample temperature.