View stored data, View stored probe data, Print stored data – Hach-Lange HQD Portable Meter User Manual User Manual

Page 16

Advertising
background image

View stored data

The data log contains sample, calibration and check standard data. The
most recent data point in the data log is tagged as Data Point 001.

1. Push the

key.

2. Select View Data Log to view the stored data. The most recent data

point is shown. The top of the screen shows whether the data is from
a sample reading, a calibration or a check standard. Push the
key to view the next most recent data point.

Option

Description

Reading Log

Reading Log—shows sample measurements including
the time, date, operator and sample ID. Select Details to
view the associated calibration data.

Calibration Log

Calibration Log—shows calibration data. Select Details
to view additional information about the calibration.

Check Standard
Log

Check Standard Log—shows check standard
measurements. Select Details to view the calibration
data that was associated with the measurement.

View stored probe data

Make sure that a probe is connected to the meter. If two probes are
connected, select the appropriate probe when prompted.

1. To view the calibration data that is stored in a probe, push the

key and select View Probe Data. The current calibration and
calibration history for the probe can be viewed.

Option

Description

View Current
Calibration

The current calibration information shows the calibration
details for the most recent calibration. If the probe has
not been calibrated by the user, the factory calibration
data is shown.

View Calibration
History

The calibration history shows a list of the times when the
probe was calibrated. Select a date and time to view a
summary of the calibration data.

Print stored data

The meter must connect to AC power to start the USB connection. Make
sure that the connection to AC power is made before the meter is
powered on.
All data can be sent to a printer. The PD-24 Citizen Printer is compatible
with the HQd meters and is FCC Part 15B, Class B compliant with the
HQ meters. Other printers may not be compatible. Compatible printers
should support a minimum of 72 columns of data, be capable of printing
up to 500 continuous data-stream events in 1, 2 and 3 lines of text and
fully support code page 437 and code page 850.

1. Turn off the meter. Make sure that the meter is connected to AC

power. Refer to

Connect to AC power

on page 8.

2. Connect the printer to the meter with a USB cable type A. Refer to

Figure 10

.

3. Turn on the meter.
4. Push the

key.

5. Select Send Data Log. Wait for the display to show “Transfer

Complete” and wait for the printer to stop printing. Disconnect the
printer.

14 English