Calibration reports
Calibration data is printed when the data log is sent to the printer or when probe data is sent to the printer. Calibration data is printed with 2 lines of
information (

Figure 17

).

Figure 17 Calibration report

1 Report type (CL = calibration, IC = current calibration)

7 Probe serial number (a “<“ or “>” on the HQ440d meter indicates the probe

position)

2 Parameter (pH, LDO, etc.)

8 Calibration slope/ratio/constant

3 Method name for probe settings

9 Offset—contents vary depending on type of parameter and user settings. May

be blank.

4 Time of calibration, prefaced by “CAL” and displayed as hh:mm in 24 h (or user-

defined) format

10 r²: contents vary depending on type of parameter being read, configuration of

specific method and number of calibration standards used. May be blank.

5 Probe model

11 Operator ID: user-defined; shows “- - -” if undefined

6 Date of calibration (DD-MM-YY or user-defined format)

The calibration history can be printed from the probe data menu. Refer to

Figure 18

.