Table 3 Operator access options

Option

Description

(Probe model) settings Only methods (if methods exist) can be selected. Refer

to the probe documentation.

Note: A probe must be connected to the meter.

Run check standard

Measure standard solution (available for pH,
conductivity, ORP and ISE probes)

Note: A probe must be connected to use this option.

Instrument information Probe information

Meter information

Access password

Enter password

Display options

Contrast

Auto shutoff

Backlight

Mode

Sounds

Key press

Stability alert

Calibration reminder

Set the display options

Use Display Options to change the display contrast, battery saving auto-
shutoff options, the backlight option or the detailed or big reading screen
mode.

1. Push the key and select Display Options.
2. Select which display option to change.

Option

Description

Contrast

Adjust the contrast of the display. The lightest setting is 0 and
the darkest setting is 9.

Option

Description

Auto-
shutoff

To maximize battery life, set a time period after which the
meter will automatically power off if no key is pushed (1, 2, 5,
10, 30 min, 1 h, 2 h or never). Auto-shutoff is not active when
the meter is connected to AC power or in the Interval Reading
Mode.

Backlight

The display backlight is turned off when the

key is pushed.

Is it possible to set a time period after which the backlight will
automatically power off if no key is pushed.

Mode

Select Detailed or Big screen size. Detailed will show more
information with smaller numbers and text. Big will show less
information with larger numbers and text.
Note: The screen size can also be selected from the measure
mode (refer to

Display description

on page 9).

Set the sound options

The meter can make an audible sound when a key is pushed, when
stability is reached or when the calibration reminder is due. The meter
also makes an audible sound when it begins transferring data to a USB
storage device and again when the data transfer is complete.

1. Push the key and select Sound.
2. Choose which events will produce an audible sound. Multiple items

can be selected.

Option

Description

Key Press

The meter will make an audible sound whenever a key is
pushed.

Stability Alert The meter will make an audible sound whenever

measurement stability is reached.

Cal reminder

The meter will make an audible sound when calibration is
due.
Note: Refer to

Set the probe calibration reminder

to set the

calibration reminder to on or off.

20 English