Message or symptom

Possible cause

Action

Connect a probe or probe
requires service

Probe not connected properly

Disconnect, then connect the probe. Tighten the locking
nut.

Software not updated

To download the most current version of the software,
refer to the applicable product page on the manufacturer's
website.
Refer to the HQd Series meter manual.

Large number of methods stored

Continue to let probe connect. Do not disconnect the
probe.

Damaged probe

Make sure there is connectivity with another probe or
meter to confirm isolated issue with probe. Contact a
Technical Support Representative.

Standard not recognized
error

Tape not removed from the
reference junctions

Remove the tape.

Standard not recognized
error

Probe storage cap or soaker
bottle not removed.

Remove the probe storage cap or soaker bottle.

Incorrect or contaminated buffer
solution

Use fresh buffer solution as specified in the method.

pH reading and/or mV
reading is same for all
solutions

Tape not removed from the
reference junctions

Remove the tape.

pH reading and/or mV
reading is same for all
solutions

Storage cap or soaker bottle not
removed

Remove the storage cap or soaker bottle.

Electrical issue

Contact a Technical Support Representative.

Slow stabilization time

Tape not removed from the
reference junctions

Remove the tape.

Slow stabilization time

Contaminated glass sensor

Clean the probe (refer to

Clean the probe

on page 9).

Poor contact between reference
junction and solution

Shake the probe in the solution from side to side to refresh
the reference junction.

Probe not prepared for sample

For the best stabilization time, soak the probe in the
sample for 10 to 15 minutes before doing a sample
measurement.

Low sample temperature or
temperature difference between
samples

Check the sample temperature. The lower the
temperature or the greater the difference of temperatures
between samples, the longer the stabilization time will be.

Air bubbles around inner
reference electrode

Gently tap the probe with hand or shake the probe
downward to remove any air bubbles.

Calibration errors

Calibration not done correctly

Recalibrate using freshly prepared pH buffers.

Contaminated glass sensor

Clean the probe (refer to

Clean the probe

on page 9).

Slope exceeds the criteria for %
of theoretical (as defined in the
method slope limit)

1.

Widen the slope limits by changing the
PHC101 calibration settings and method.

2.

Recalibrate the probe.

3.

Run a check standard to check the performance of
the probe.