Message or symptom

Possible cause

Action

Drifting/Inaccurate readings

Contaminated glass sensor

Clean the probe (refer to

Clean the probe

on page 9).

CO

2

absorption (for low ionic

strength or high purity sample)

Use the LIS chamber for LIS/high purity samples to keep
sample contamination from occuring.

Air bubbles around inner
reference electrode

Rinse the reference junction holes with deionized water
and then gently tap the probe with hand or shake the
probe to remove any air bubbles.

Incorrect storage conditions
(discolored, contaminated or
dried gel)

Clean or condition the probe and do recalibration. The
probe may not operate correctly if the probe has been left
dry for a long time. Condition the glass sensor and
reference junctions again:

1.

Soak the probe tip in the 4.01, 7.00 and 10.01 buffers
each for 5 minutes.

2.

Rinse with deionized water. Blot dry with a lint-free
cloth.

3.

Calibrate the probe.

Electromagnetic Forces (EMF)
such as voltaic cells,
thermoelectric devices, eletrical
generators, resistors and
transformers.

Do not test in areas where EMF is present. For testing in
process units (i.e. spot checking), make sure the
equipment is grounded.

Out of range

Measurement value is outside of
range

Make sure that the sample is within the range of the
probe.

Temperature out of range

Temperature value is outside of
range

Make sure that the sample temperature is within the range
of the probe.

Make sure that the temperature sensor is working
correctly.

Measured pH buffer temperature
is outside range of the probe

Make sure that the standard temperature is within the
range of the probe.

Make sure that the temperature sensor is working
correctly.

Check standard temperature
value is outside of range

Make sure that the Check Standard temperature is within
the range of the probe.

Out of limits

Measurement value is outside of
measurement limits set in the
current method

Make sure that the sample is within the limits of the
current method.

Create a new method with expanded limits.

Check standard value is outside
of limits set in the current method

Make sure that the check standard is within the limits of
the current method.

Create another method that expands the acceptable limits.

HACH COMPANY World Headquarters
P.O. Box 389, Loveland, CO 80539-0389 U.S.A.
Tel. (970) 669-3050
(800) 227-4224 (U.S.A. only)
Fax (970) 669-2932
[email protected]
www.hach.com

HACH LANGE GMBH
Willstätterstraße 11
D-40549 Düsseldorf, Germany
Tel. +49 (0) 2 11 52 88-320
Fax +49 (0) 2 11 52 88-210
[email protected]
www.hach-lange.de

HACH LANGE Sàrl
6, route de Compois
1222 Vésenaz
SWITZERLAND
Tel. +41 22 594 6400
Fax +41 22 594 6499

©

Hach Company/Hach Lange GmbH, 2010, 2013. All rights reserved. Printed in U.S.A.

05/2013, Edition 2

Advertising