Advanced operation, Change measurement options – Hach-Lange PHC10115 User Manual User Manual

Page 6

Advertising
background image

7. If the display shows Check Standard Passed, the check standard measurement is

within the accepted limits set by the administrative user. Select Done to continue with
the sample measurement.

8. If the display shows Check Standard Failed, the measurement is outside of

accepted limits set by the administrative user and a recalibration is recommended. If
the acceptance criteria is set to Cal Expires on Failure: Yes, the display shows the
calibration icon and a question mark until the probe is recalibrated. To correct the
probe calibration and status indicator, calibrate the probe (refer to

Calibration

on page 3).

Advanced operation

Parameter-specific settings can be changed through the Full Access Options menu.
Details about menu navigation, available options and how to change them are given in
the screens, tables and procedures throughout this section.

The settings that can be changed are shown in

Table 1

.

Table 1 Parameter-specific settings

Setting

Options

Measurement Options

Resolution

Upper and lower range limits

Calibration Options

Buffer set

Buffer set values (if Custom Buffer Set selected)

Reminder

Minimum Cal points

Slope limit

Check Standard Options

Standard (temperature compensated buffer or custom at 25 °C)

Reminder

Acceptance criteria

Standard value at 25 °C (if custom standard selected)

Units

pH

mV

Change measurement options

Methods are groups of default or user-defined settings relevant to specific applications. If
the meter is set to the default method and the Modify Current Settings option is chosen, a
prompt for a new name is shown after the changes are entered. The settings are saved
with this name to distinguish them from the default method settings, which cannot be
changed. A saved method can be used instead of multiple adjustments to the individual
settings. Changes made to a user-defined method are automatically saved with the
existing name. Multiple methods can be saved for the same probe on each meter.

1. Make sure a probe is connected to the meter.
2. Push and select PHC101 Settings.

6