Run a check standard – Hach-Lange PHC10115 User Manual User Manual

Page 5

Advertising
background image

Note: Procedure also applies for rugged model probes.

1. Connect the
probe to the
meter. Make sure
that the cable
locking nut is
securely
connected to the
meter. Turn the
meter on.

2. For the best
stabilization time,
condition the
probe for several
minutes in the
sample or in a
solution
comparable to the
sample in terms of
pH and ionic
strength before
the initial sample
measurement.

3. Rinse the
probe with
deionized water,
then with the
sample. Blot dry
with a lint-free
cloth.

4. Put the probe
in the sample and
stir gently. Make
sure that the
reference
junctions are
completely
submerged. Do
not put the probe
on the bottom or
sides of the
container. Shake
the probe from
side to side in the
sample to refresh
the reference
junction.

5. Push Read.
The display will
show "Stabilizing"
and a progress
bar as the probe
stabilizes in the
sample. The
display will show
the lock icon when
the reading
stabilizes.

6. Repeat steps
3-5 for additional
measurements.

7. When
measurements
are done, store
the probe (refer to

Storage

on page 10).

Run a check standard

The run check standard feature validates instrument performance between sample
measurements. Use the run check standard feature for periodic or user-defined interval
measurements of a traceable standard solution. Set the criteria for check standards from
the PHC101 Settings menu.

Note: Access control must be off or a valid password must be entered before any of the check

standard method options can be changed.

1. Push . The Full Access Options menu is shown.
2. Select Run Check Standard.

Note: Select the correct probe if two probes are connected to the meter.

3. Get the standard solution shown on the display.
4. Rinse the probe with deionized water. Blot dry with a lint-free cloth.
5. Put the probe in the standard solution until the temperature sensor is completely

submerged. Move the probe up or down or gently tap on the beaker to remove air
bubbles from the probe.

6. Push Read. The display will show "Stabilizing" and a progress bar as the reading

stabilizes. The display shows the value of the check standard and either Check
Standard Passed or Check Standard Failed.

5