Troubleshooting, Replacement parts and accessories, Clean the probe as needed. refer to – Hach-Lange SENSION+ EC71 User Manual

Page 14

Advertising
background image

Troubleshooting

Refer to the following table for common problem messages or
symptoms, possible causes and corrective actions.

Table 2 Calibration warnings and errors

Error/Warning

Solution

UNSTABLE
READING
Time t> 100 s

Calibrate again.
Examine the probe: Make sure that the probe is properly
immersed in the sample; make sure that there are no air
bubbles inside the probe measuring chamber; rinse the
probe with ethanol.

Check temperature.
Check the cell.

Make sure the standard temperature is constant during
the calibration.
Examine the probe: Make sure that the probe is properly
immersed in the sample; make sure that there are no air
bubbles inside the probe measuring chamber; rinse the
probe with ethanol.

STANDARD
TEMPERATURE
< 15 °C or > 35 °C

Calibrate again.
The standard temperature must be between 15 and
35 ºC.

Difference C > 30 %

Calibrate again.
Examine the probe: Make sure that the probe is properly
immersed in the sample; make sure that there are no air
bubbles inside the probe measuring chamber; rinse the
probe with ethanol; connect a different probe to verify if
problem is with probe or meter.
Examine the standard solution: Use new standard
solution.

C < 0.05 or > C >
50.00 °C

SAME BUFFERS

Buffer or cell in poor
conditions

BUFFER NOT
RECOGNIZED

Difference C > 30 %

SAME STANDARDS

Table 3 Measurement warnings and errors

Error/Warning

Solution

23.2 °C
728 µS/cm @ 25 °C

Calibrate again.

EC out of range

Examine the probe: Make sure that the probe is
properly immersed in the sample; make sure that
there are no air bubbles inside the probe measuring
chamber; rinse the probe with ethanol; connect a
different probe to verify if problem is with probe or
meter.

Out of range °C

Examine the temperature sensor.
Connect a different probe to verify if problem is with
probe or meter.

TC = 0
Impossible to measure
salinity

Modify TC

Time > 150 s

Examine the temperature.
Examine the probe: Make sure that the probe is
properly immersed in the sample; make sure that
there are no air bubbles inside the probe measuring
chamber; rinse the probe with ethanol; connect a
different probe to verify if problem is with probe or
meter.

Replacement parts and accessories

Note: Product and Article numbers may vary for some selling regions. Contact the
appropriate distributor or refer to the company website for contact information.