Change the language, Standard operation, Calibration – Hach-Lange SENSION+ EC71 User Manual

Page 8: Calibration settings, Calibration types, Calibration procedure

Advertising
background image

Change the language

The display language is selected when the meter is powered on for the
first time.

1. Use the or to select a language from a list.
2. Confirm with . The measurement screen shows DATA OUTPUT.
3. Select Deactivated if no printer or PC is connected and confirm.

Refer to

Select the data output

on page 9 for more information

about Data Output.

Standard operation

Calibration

W A R N I N G

Chemical exposure hazard. Obey laboratory safety procedures and wear all of
the personal protective equipment appropriate to the chemicals that are handled.
Refer to the current material safety data sheets (MSDS) for safety protocols.

Calibration settings

The calibration settings contain Calibration type, Calibration frequency
and Display options.

1. From the main menu use the or to select CALIBRATION.

Confirm.

2. Use the to enter the calibration menu.
3. Use the or to select the following options:

Option

Description

Calibration type Calibration type—select Molar Standards, Demal

Standards, NaCl Standards, Calibration to a X value, data
introduction or Theoretical calibration. Refer to

Calibration

types

for more information.

Option

Description

Cal. frequency

Calibration reminder—can be set between 0–99 days
(default 15 days). The display shows the remaining time
to the new calibration. Refer to

Set the calibration

reminder

on page 9 for more information.

Display Ω

Display Resistivity—select YES to show simultaneously
conductivity and resistivity.

Calibration types
Different calibration types can be selected.

1. From the main menu use the or to select CALIBRATION.

Confirm.

2. Use the to enter the calibration menu.
3. Use the or to select Calibration type.

Option

Description

Molar Standards

147 µS/cm, 1413 µS/cm, 12.88 mS/cm and
111.8 mS/cm at 25 °C (77 °F)

Demal Standards

1049 µS/cm, 12.85 mS/cm and 111.31 mS/cm at
25 °C (77 °F)

NaCl Standards

1014.9 µS/cm 25 °C (77 °F)

Calibration to a X value To adjust manually any scale value of the

measured conductivity.

Data introduction

Manual probe constant introduction.

Theoretical calibration

The probe calibration data is replaced with
C=1.000 cm

-1

.

Calibration procedure

This procedure is for general use with liquid calibration solutions. Refer
to the documents that are included with each probe for additional
information.

Note: Solutions should be stirred during calibration. For more information about the
stirring settings, refer to

Change the stirring settings

on page 10.

8 English

Advertising