View the calibration data, Set the calibration reminder, Sample measurements – Hach-Lange SENSION+ pH31 User Manual

Page 10: Advanced operation, Use a sample id, The new calibration. refer to

Advertising
background image

View the calibration data

Data from the most recent calibration can be shown.

1. From the main menu use the or to select DATA LOGGER.

Confirm.

2. Select Display data.
3. Select Calibration data and confirm with . The last calibration data

is shown.

• pH-the slope and offset values are shown alternating with the

deviation (in %) and calibration temperature.

• ORP-the measured mV value and calibration temperature are

shown.

• Conductivity-the cell constant and calibration temperature for each

standard are shown.

Set the calibration reminder

The calibration reminder can be set between 0 to 23 hours or 1-7 days
(default 1 day). The display shows the remaining time to the new
calibration.

Note: When 0 days is selected, the calibration reminder is turned off.

1. From the main menu use the or to select CALIBRATION.

Confirm.

2. Use the to enter the calibration menu.
3. Use the or to select Cal. frequency and confirm.
4. Use the and to advance to the next step and use the or to

change a value. Confirm.
Push to start the calibration.

Sample measurements

Each probe has specific preparation steps and procedures for taking
sample measurements.

1. From the main menu use the or and and to select

MEASURE. Confirm.

2. Use the to change the following settings. Confirm every entry.

Option

Description

Resolution

Select the resolution: 1, 0.1, 0.01 (default) or 0.001

Measure

Stability—select By stability Criterion: Fast (variation <
0.02 pH in 6 s), Standard (variation < 0.01 pH in 6 s) or Strict
(variation < 0.002 pH in 6 s). In continuous—enter the time
interval for the In continuous Acquis. interval (data storage or
printing data). By time—enter the time interval for data storage
or printing data.

Display mV Display mV—select YES or NO to show mV.

Limits

Limits—select YES or NO. YES: Enter the upper and lower
limits. An acoustic warning appears when the measurement is
out of limit. The report output shows an A next to the
measured value when the measurement was out of limit.

Isopotential Isopotential—change the Isopotential pH value in Data

introduction. Select Calculate to calibrate the probe again.

3. Push to start the measurement.

Note: If the measurement is not stabilizing after 120 seconds, the meter turns
automatically into the continuous measurement mode.

Advanced operation

Use a sample ID

The sample ID tag is used to associate readings with a particular sample
location. If assigned, stored data will include this ID.

1. From the main menu use the or to select SYSTEM. Confirm.
2. Use the or to select Sample ID and confirm.

10 English