Troubleshooting, Clean the probe as needed. refer to – Hach-Lange SENSION+ pH31 User Manual

Page 16

Advertising
background image

Troubleshooting

Refer to the following table for common problem messages or
symptoms, possible causes and corrective actions.

Table 2 Calibration warnings and errors

Error/Warning

Solution

Asymmetry >
58 mV

Slope out of range (accepted values ± 58 mV).
Calibrate again.
Connect a new probe.

Buffer or electrode
in poor conditions.

Calibrate again.
Examine the probe: Clean the probe (refer to

Clean the

probe

on page 13 for more information); make sure that

there are no air bubbles in the membrane. Shake the probe
like a thermometer; Connect a different probe to verify if
problem is with probe or meter.
Examine the buffer solution: Make sure that the buffer used
matches the buffer specified in configuration; make sure of
the temperature specification in configuration; use new
buffer solution.

Sens. (a) < 70%

BUFFER NOT
RECOGNIZED

UNSTABLE
READING
Time t> 100 s

Calibrate again.
Examine the probe: Clean the probe (refer to

Clean the

probe

on page 13 for more information); make sure that

there are no air bubbles in the membrane. Shake the probe
like a thermometer; connect a different probe to verify if
problem is with probe or meter.
Make sure that the membrane and the diaphragm are
properly immersed in the sample.

Table 2 Calibration warnings and errors (continued)

Error/Warning

Solution

Electrode in poor
conditions.

Examine the probe: Clean the probe (refer to

Clean the

probe

on page 13 for more information); make sure that

there are no air bubbles in the membrane. Shake the probe
like a thermometer; connect a different probe to verify if
problem is with probe or meter.

Check the
electrode

SAME BUFFERS

Calibrate again.
Examine the probe: Clean the probe (refer to

Clean the

probe

on page 13 for more information); make sure that

there are no air bubbles in the membrane. Shake the probe
like a thermometer; connect a different probe to verify if
problem is with probe or meter.
Examine the buffer solution: Use new buffer solution.

Table 3 Measurement warnings and errors

Error/Warning

Solution

pH 12.78 19°C

Calibrate again

pH out of range

Examine the probe: Clean the probe (refer to

Clean the

probe

on page 13 for more information); make sure that

there are no air bubbles in the membrane. Shake the
probe like a thermometer; connect a different probe to
verify if problem is with probe or meter.

Temp out of range
°C

Examine the temperature sensor.
Connect a different probe to verify if problem is with probe
or meter.

Time > 60 s

Make sure that the membrane and the diaphragm are
properly immersed in the sample.
Examine the temperature.
Examine the probe: Clean the probe (refer to

Clean the

probe

on page 13 for more information); make sure that

there are no air bubbles in the membrane. Shake the
probe like a thermometer; connect a different probe to
verify if problem is with probe or meter.

Time > 150 s

Time > 300 s

16 English