Standard solutions, Are not used. refer to, Technical buffer solutions – Hach-Lange SENSION+ pH31 User Manual

Page 18: Din 19267), Replacement parts and accessories (continued)

Advertising
background image

Replacement parts and accessories (continued)

Description

Item no.

3x50 mL printed beakers for benchtop pH calibration

LZW9110.98

Three-sensor holder, for sensION+ benchtop instruments

LZW9321.99

Holder and clamp for three sensors

LZW9155.99

Pyrex glass chamber, continuous flow measurements

LZW9118.99

PP protector, electrode storage

LZW9161.99

Standard solutions

Technical buffer solutions (DIN 19267)
Refer to

Table 4

pH and ORP (mV) values of specific buffer sets at

varying temperatures.

Table 4 pH, ORP (mV) and temperature values

Temperature

pH

mV

°C

°F

0

32

2.01

4.01

7.12

9.52

10.30

10

50

2.01

4.00

7.06

9.38

10.17

245

20

68

2.00

4.00

7.02

9.26

10.06

228

25

77

2.00

4.01

7.00

9.21

10.01

220

30

86

2.00

4.01

6.99

9.16

9.96

212

40

104

2.00

4.03

6.97

9.06

9.88

195

50

122

2.00

4.06

6.97

8.99

9.82

178

60

140

2.00

4.10

6.98

8.93

9.76

160

70

158

2.01

4.16

7.00

8.88

80

176

2.01

4.22

7.04

8.83

90

194

2.01

4.30

7.09

8.79

18 English

Advertising