Clean the probes – Hach-Lange SENSION+ 50 49 User Manual User Manual

Page 13

Advertising
background image

Pull straight to remove or connect the probes. Do not turn the connector. Look to the color coding of
the probe and multisensor connectors. Use the white mark to position the probes in the multisensor
connectors. Refer to

Figure 2

for more details.

Figure 2 Replace the probes

1 Blue connector: Dissolved

oxygen probe

2 Green connector: Conductivity

probe

3 Black connector: pH probe

Clean the probes

N O T I C E

Do not rub or touch the tip of the pH or dissolved oxygen sensors.

Clean the probe when there is contamination on the sensor. Symptoms of contamination are:

• Readings are not accurate or consistent.
• The stabilization time is slow.
• A calibration error occurs.
• Contamination is visible on the probe (i.e., dirt).

To clean the dissolved oxygen probe of the multisensor, refer to

Dissolved oxygen probe

on page 12.

1. For general contaminants, rinse the probe with deionized water. Blot dry with a lint-free cloth.
2. For other contaminants:

a. Soak the pH and/or conductivity probes of the multisensor in the applicable cleaning agent.

Refer to

Table 2

.

b. Rinse or soak the sensors for 1 minute in deionized water. Blot dry with a lint-free cloth.

English 11