5. If the probe will be used immediately, polarize the probe. Refer to

Polarize the dissolved oxygen

probe

on page 5.

6. If the probe will not be used immediately, put the probe in storage. Refer to

Storage

on page 13.

Figure 3 Fill the membrane module

1 Anode

2 Cathode

3 Membrane module

Storage

1. Fill the probe storage tube with storage solution to the mark. Refer to

Specifications

on page 3 for

the applicable storage solution.

2. Put the probe in the storage tube and turn to tighten.

Rehydrate the pH probe

N O T I C E

Do not clean the dissolved oxygen probe of the multisensor with HCl because damage to the membrane will
occur.

If the pH sensor of a probe becomes dry, rehydrate the probe. A dry pH probe will not operate
correctly.

1. Soak the dry probe in a dilute HCl solution for several hours.
2. Rinse the probe with deionized water. Blot dry with a lint-free cloth.
3. Calibrate the probe. Refer to

Calibration procedure (pH and conductivity)

on page 6.

Troubleshooting

Refer to the table that follows for common problem messages or symptoms, possible causes and
corrective actions.