Troubleshooting notes, Examine the probes, Replacement parts and accessories – Hach-Lange SENSION+ 50 49 User Manual User Manual

Page 17: Troubleshooting notes examine the probes

Advertising
background image

Error/Warning

Description

Solution

E6

The calibration solutions
have different
temperatures.

Make sure that the calibration solutions have the same
temperature.

ErA

Salinity is set to automatic
and measured value >
50 g/L.

Change the TC value from automatic to the correct value. Refer to
meter manual.

Troubleshooting notes

For the best performance, make sure to:

• Prepare the probe(s) for use before calibration or measurement.
• Obey the calibration and measurement notes.
• Obey the calibration and measurement procedures.

Troubleshooting notes:

• Magnetic stirrers may supply sufficient heat to change solution temperature. Put a piece of

insulating material between the stirrer and beaker.

• Protect the sensing element from direct light during measurement.
• Set the meter to off. Disconnect and then connect the probe again.
• Connect a different probe to identify if the problem is with the probe or the meter.

Refer to

Troubleshooting

on page 13 for calibration and measurement warnings and errors.

Examine the probes

Note: The lower the sample temperature or the larger the temperature difference between the samples, the longer
the stabilization time will be.

1. Examine the platinized layer of the conductivity probe. Replace the probe if the platinized layer

has been damaged and/or removed.

2. Clean the probes. Replace the DO probe membrane module if necessary.
3. If the DO probe fill solution is contaminated, drain and replace with fresh filling solution.
4. Rinse the pH probe reference junction diaphragm with deionized water. Then, shake the probe

downwards (like a thermometer) to remove any air bubbles in the bulb.

Replacement parts and accessories

Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.

Replacement parts

Description

Item no.

pH electrode for 50 49 multisensor

LZW9171.99

Conductivity cell for 50 49 multisensor

LZW9172.99

DO sensor for 50 49 multisensor

LZW9173.99

Handle for 50 49 multisensor

LZW9179.99