Operation – Hach-Lange LZV902 User Manual

Page 45

Advertising
background image

English 45

Operation

Measurements with the carousel insert

Stored programs

1.

Select Stored Programs in the main menu.

2.

Select the desired test and press Start.

3.

Press Options > More > Reading mode.

4.

Select Carousel 1 inch square.

5.

In the toolbar, select the symbol for the carousel insert. This symbol

opens a window with carousel options; refer to

Table 1

.

6.

Equip the first carousel insert with zero solution and samples, as

per the cell order information. The zero solution is placed in

position 1.

Table 1 "Stored Programs" carousel options

Options

Description

Number of samples

Number of samples without zero solution (or blank
value), maximum 40 samples possible.

Sample ID

Off: The samples are not assigned a sample ID
Enter Sample ID: The sample is selected and a sample
is entered under Edit. The sample IDs are not saved in
the general sample ID list, rather they only apply to the
current series of measurements.
Select Sample ID: The sample is selected and the
general sample ID list opens under Select. Each
sample is allocated the desired sample ID.
Import external Sample ID list: *.txt and *.csv files can
be imported via a USB stick or network; refer to

section

Import an external sample ID

.

Cell Order Information

Cell sequence indicator: shows the position in the
carousel insert in which each sample must be placed.
Z1 is the zero solution (blank value). Defined sample
IDs are specified here and make assigning the samples
easier.

Dilution Factor

Each sample can be assigned a dilution factor.

Table 1 "Stored Programs" carousel options

Options

Description

Advertising