English 47

5.

Load the carousel insert according to the selected carousel option.

6.

Start the measurement by pressing Zero or Read or Zero & Read

according to the selected carousel option.
The carousel insert is measured and the results are displayed.

Note: If Instrument Setup > Data Log Setup > Auto Store: On has
been selected, all measurement data is automatically saved. The
measurement data can be accessed under Recall data > Data Log.

Multi-wavelength

1.

Press Multi-Wavelength in the main menu.

2.

Press Options > More > Reading mode.

3.

Select Carousel 1 inch square or Carousel 10 mm square.

4.

In the toolbar, select the symbol for the carousel insert. This symbol

opens a window with carousel options; refer to

Table 2

.

5.

Load the carousel insert according to the selected carousel option.

6.

Start the measurement by pressing Zero or Read or Zero & Read

according to the selected carousel option.
Each zero solution and sample is measured with the defined
wavelength and the results are displayed.

Note: If Instrument Setup > Data Log Setup > Auto Store: On has
been selected, all measurement data is automatically saved. The
measurement data can be accessed under Recall data > Data Log.

Wavelength scan

1.

Press Wavelength Scan in the main menu.

2.

Press Options > More > Reading mode.

3.

Select Carousel 1 inch square or Carousel 10 mm square.

4.

In the toolbar, select the symbol for the carousel insert. This symbol

opens a window with carousel options; refer to

Table 2

.

5.

Load the carousel insert according to the selected carousel option.

6.

Start the measurement by pressing Zero or Read or Zero & Read

according to the selected carousel option.
Each zero solution and sample is measured with the defined
wavelength range; the curve is displayed.

Save the wavelength scan

1.

Press Options > Select View and select the cell for which you

want to save the scan.

2.

Select Table or Graph; the scan is displayed.

3.

Press Options > Store Symbol and select one of the 20 scan

storage locations.

4.

Press Store.

5.

The measurement data can be accessed in the main menu under

Recall Data > Wavelength Scan.