Accessories – Hach-Lange LZV902 User Manual

Page 48

Advertising
background image

English 48

Time course

1.

Press the Time Course menu option in the main menu.

2.

Press Options > More > Reading mode.

3.

Select Carousel 1 inch square or Carousel 10 mm square.

4.

In the toolbar, select the symbol for the carousel insert. This symbol

opens a window with carousel options; refer to

Table 2

.

5.

Load the carousel insert according to the selected carousel option.

6.

The measurement interval between the individual measurements

must be at least 15 seconds.

7.

Start the measurement by pressing Zero or Read or Zero & Read

according to the selected carousel option.
Each zero solution and sample is measured in the defined
measurement interval and the curve is displayed.

Storing the time course

1.

Press Options > Select View and select the cell for which you

want to store the elapsed time.

2.

Select Table or Graph; the elapsed time is displayed.

3.

Press Options > Store Symbol and select one of the 20 cycle

storage locations.

4.

Press Store.

5.

The measurement data can be accessed in the main menu under

Recall Data > Time Course.

Accessories

Description

Order no.

Carousel insert (sample changer) 10 mm

LZV902.99.00001

Carousel insert (sample changer) 1 inch

LZV902.99.00011

1-inch square cells, 8 matched cells, glass (10 mL)

2495408

10 mm square cells, matched pair, glass (3.5 mL)

2095100

10 mm square cells, quartz

2624410

10 mm disposable square micro cells (1.5 mL)

2629500

10 mm square plastic cells

2743400

Plug, neoprene for 1-inch square cells

1480801