1 sensor connection and wiring, Installation 3.4.1 sensor connection and wiring – Hach-Lange UVAS sc User Manual
Page 16
Advertising
16
Installation
3.4.1 Sensor connection and wiring
The sensor cable is supplied with a keyed quick-connect fitting for easy
attachment to the controller. Retain the connector cap to seal the connector
opening in case the sensor must be removed. Optional extension cables may be
purchased to extend the sensor cable length.
Figure 6
Attaching the Sensor using the Quick-connect Fitting
Figure 7
Quick-connect Fitting pin assignment
C A U T I O N
Before power is applied, refer to the controller operation instructions.
Number
Designation
Wire Color
1
+12 VDC
Brown
2
Circuit Common
Black
3
Data (+)
Blue
4
Data (–)
White
5
Shield
Shield (grey wire in existing quick-disconnect fitting)
6
Groove
1
2
3
4
5
6