2 organic loading (sac), 1 procedure organic loading (sac), 1 executing the test with the sipper module – Hach-Lange DR 5000 drinking water analysis, supplementary software LZV571 User Manual

Page 11: Procedure organic loading (sac)

Advertising
background image

11

Working procedures

2.2 Organic loading (SAC)

Principle

The parameter "organic loading” can be determined from a sample
by measuring the intrinsic absorbance, that is without using
reagents. The spectral absorbance coefficient is determined in m

-1

.

Range of application

Drinking water, surface water, wastewater

Measuring range

0–20 m

-1

Accessories

Spectrophotometer (254 nm)

50 mm rectangular cuvette (QS)

Sipper module: 50 mm pour-through cuvette (QS)

2.2.1

Procedure Organic loading (SAC)

1.

Select

Stored Programs

in the "Main Menu". Select test

number

902

.

2.

Insert the zero cuvette (distilled water) in the cell compartment.
Close the cell compartment. Touch

Zero

.

3.

Insert the sample cuvette containing the first sample in the cell
compartment and close the cell compartment. Touch

Read

.

The result is displayed.

Note:

Analysis of additional samples: Repeat working procedure from

point 3.

2.2.1.1 Executing the test with the sipper module

Information about the installation, module configurations and
sample introduction of the sipper module can be found in the user
manual of the DR 5000 Spectrophotometer (15.3.3. Installation
Sipper Module, page 133).