2.1 Nitrate (intrinsic absorbance)

Principle

The parameter "nitrate” can be determined in a water sample by
measuring the intrinsic absorbance, that is without using reagents.
The nitrate ion is determined bichromatically at two wavelengths.
The difference between the measured absorbance values at the
two wavelengths is a measure of the nitrate concentration, which
can be evaluated relative to a standard concentration.

Range of application

Drinking water

Measuring range

1–60 mg/l nitrate (3 mm rectangular cuvette)
0.3–15.0 mg/l nitrate (10 mm rectangular cuvette)

Accessories

Spectrophotometer (228 and 218 nm)

3 mm or 10 mm rectangular cuvette (QS)

Sipper module: 3 mm or 10 mm pour-through cuvette (QS)

Reagents

Standard 30 mg/l nitrate for 3 mm rectangular cuvette

Standard 10 mg/l nitrate for 10 mm rectangular cuvette

Interferences

Organic components can interfere with the determination. Very high
nitrate concentrations of several 100 mg/l cause results within the
measuring range to be obtained.

Literature

E. Goldman, R. Jacobs:
Determination of Nitrates by Ultraviolet Absorbance, Amer. Water
Works Assoc. 53, 187–191, (1961)

2.1.1

Procedure Nitrate (intrinsic absorbance

)

1.

Select

Stored Programs

in the "Main Menu". Select test

number

901

(measuring range 3 mm or 10 mm).

2.

Insert the zero cuvette (distilled water) in the cell compartment.
Close the cell compartment. Touch

Zero

.

3.

Insert the standard cuvette containing the nitrate standard
(30 mg/l – 3 mm rectangular cuvette – or 10 mg/l – 10 mm
rectangular cuvette) in the cell compartment and close the cell
compartment. Touch

Read

. Display:

E1

and

E2

.

4.

Insert the sample cuvette containing the first sample in the cell
compartment and close the cell compartment. Touch

Read

.

Display:

E3

. The result is then displayed.