2 data table, 3 programming procedure, 1 lck 310 chlorine cl2 – Hach-Lange PCII-Chlorine LCK 310 User Manual

Page 23: 1 lck 310 chlorine cl

Advertising
background image

System start up

22

LCK 310: 0.05-2 mg/l Cl

2

or 0.09-3.8 mg/l ClO

2

or 0.05-2 mg/l O

3

Select the calibration data of maximum two measuring ranges which will be the

suitable one.

Each programming procedure has only to be followed once prior to the first test

evaluation.

For detailed information please read the POCKET COLORIMETER II instruction

manual.

3.2.2 Data table

3.2.2.1

Evaluation as Chlorine Cl

2

, Chlorine dioxide ClO

2

and Ozone O

3

3.3 Programming procedure

Note: Each programming procedure has only to be followed once prior to the first test
evaluation.

Important Note: The following programming procedure is valid for the evaluation

form Chlorine Cl

2

. To evaluate as Chlorine dioxide ClO

2

or Ozone O

3

please choose

the data mentioned in table

3.2.2.1 Evaluation as Chlorine Cl2, Chlorine dioxide

ClO2 and Ozone O3

.

Note: After programming the calibration data, it could be helpful to mark the channel(s) 1
and/or 2 accordingly with a label.

3.3.1 LCK 310 Chlorine Cl

2

1.

Press POWER to switch on the instrument.

2.

Press MENU. "SEL" is displayed. A flashing arrow indicates the current

measuring range (1 or 2).

3.

Select measuring range 1 or 2 for LCK 310 (0.05-2 mg/l Cl

2

).

Note: Press

GREEN

key to toggle between measuring ranges.

Table 1 Evaluation as Chlorine Cl

2

, Chlorine dioxide ClO

2

and Ozone O

3

LCK 310

Chlorine

LCK 310

Chlorine dioxide

LCK 310

Ozone

RES (resolution)

--.--

--.--

--.--

S0 (standard 0)

00.05

00.09

00.05

A0 (absorbance 0)

0.015

0.015

0.025

S1 (standard 1)

02.00

03.80

02.00

A1 (absorbance 1)

0.580

0.600

0.900

Advertising