Section 4 operation, 2 quality assurance, Mode. see – Hach-Lange PCII-Chlorine LCK 310 User Manual

Page 26

Advertising
background image

25

Section 4

Operation

4.1 Analyse Chlorine/Chlorine dioxide/Ozone with

LANGE Cuvette Tests

DANGER

Handling chemical samples, standards and reagents can be dangerous. Users

of this product are advised to familiarize themselves with safety procedures

and the correct use of chemicals, and to carefully read all relevant material

safety data sheets.

To analyse Chlorine, Chlorine dioxide or Ozone follow the working procedure

enclosed in the according Cuvette Test LCK 310.

4.2 Quality assurance

The measurement results must be subjected to plausibility checks (dilute and/or

spike the sample).

Regular checks with corresponding standard solutions (see

Section 5 Replacement

Parts and Accessories on page 27

) ensure that your measurement system is

functioning properly and is being correctly operated.

4.3 Evaluate LANGE Chlorine/Chlorine

dioxide/Ozone Cuvette Tests

Note: Make sure that the instrument is programmed with the corresponding calibration data,
see

3.4 Check entered data on page 24

.

1.

Press POWER to switch on the instrument.

2.

Make sure that the correct measuring range is selected!

Otherwise press MENU. "SEL" is displayed. A flashing arrow indicates the

current measuring range (1 or 2). Press GREEN key to select measuring range.

Press MENU to get back to the measuring mode.

3.

Insert the zero-solution into the cell holder equipped with the cuvette adapter

1 cm.

Note: Make sure that the cuvette adapter is positioned correctly.

4.

Place the instrument cap with the HACH LANGE label over the cell holder to

cover the cuvette.

5.

Press BLUE key. "0.00" is displayed.

6.

Remove the instrument cap and insert the sample cuvette.

7.

Place the instrument cap over the cell holder to cover the cuvette.

8.

Press GREEN key.

9.

The result is displayed in mg/l Chlorine/Chlorine dioxide/Ozone.

Note: In case the Chlorine/Chlorine dioxide/Ozone concentration is out of the selected
measuring range the value for the lowest or highest concentration is displayed and