Section 4 operations, 1 use of an sc controller, 2 sensor setup – Hach-Lange SOLITAX sc User Manual

Page 15: 3 sensor data logging, 4 sensor diagnostics menu for ph and orp, 5 sensor setup menu

Advertising
background image

15

Section 4

Operations

4.1 Use of an sc controller

Before using the sensor in combination with an sc controller, refer to the controller user
manual for navigation information.

4.2 Sensor Setup

When a sensor is initially installed, the serial number of the sensor will be displayed as the
sensor name. To change the sensor name refer to the following instructions:

1. Select Main Menu.

2. From the Main Menu, select SENSOR SETUP and confirm.

3. Highlight the appropriate sensor if more than one sensor is attached and confirm.

4. Select CONFIGURE and confirm.

5. Select EDIT NAME and edit the name. Confirm or cancel to return to the Sensor

Setup menu.

4.3 Sensor Data Logging

A data memory and event memory per sensor are available via the sc controller. While
measured data are saved in the data memory at stipulated intervals, the event memory
collects numerous events such as configuration changes, alarms and warning conditions.
Both the data memory and the event memory can be read out in CSV format. For
information on how you can download the data, please see the controller manual.

4.4 Sensor Diagnostics Menu for pH and ORP

SELECT SENSOR (if more than one sensor is attached)

STATUS

ERROR LIST

See

section 6.1 on page 21

.

WARNING LIST

See

section 6.2 on page 21

..

4.5 Sensor Setup Menu

SELECT SENSOR (if more than one sensor is attached)

WIPE

Initiates a wiping action on the sensor window.

CALIBRATE

SET OUTMODE

Select the behavior of the outputs during calibration for zero point setting (Hold, Active, Transfer,
Selection). Hold maintains the last reading prior to going into the menu. Active transmits the
current level readings, corrected with previous calibration data until new data is entered. Set
Transfer transmits the value designated during the system setup

SENSOR MEASURE

Displays the current, uncorrected measured value.

CONFIGURE

Select the calibration type and follow the calibration steps for 2 point, 3 point, 4 point, and 5 point
calibration.

FACTOR/2 POINTS/
3 POINTS/4 POINTS/
5 POINTS

Display depends on the selection in configuration.

SET CAL DEFAULT

Return the instrument to the default calibration settings.