Operations

4.6 Calibration

There are two calibration techniques; depending on whether turbidity or suspended solid
is required (refer to

section 4.6.2

or

section 4.6.3 on page 18

). Before calibration,

determine the behavior of the 4–20 outputs and alarm relays while the user is in the
CALIBRATE menu (refer to

section 4.6.1

).

CONFIGURE

EDIT NAME

Enter up to a 10-digit name in any combination of symbols and alpha or numeric characters.

SET PARAMETER

This setting configures the Solitax to measure turbidity or suspended solids. The Solitax cannot
simultaneously measure both. Choose “TRB” for turbidity measurements, or “TS” for suspended
solids measurement. This selection determines which units may be selected in the “Meas Units”
menu.

MEAS UNITS

Choose from the displayed units. TRB (FNU, EBC, TE/F, NTU); TS (mg/L, g/L, ppm, %)
Default: FNU
If TRB was selected in set parameter, select “NTU” (commonly used in the U.S.), FNU, EBC, or
TE/F. If TS was selected, choose mg/L, g/L, ppm, or %. Press enter to choose the selection. If
the units selected result in a reading that exceeds 4 digits, the display will only show dashes.
For example, if mg/L were selected, and the measurement was 10,500 mg/L, the display will
show dashes until the reading drops to 9999 or lower.

CLEAN INTERVAL

Select the cleaning interval (1, 5, 15 or 30 minutes; 1, 4, or 12 hours; 1, 3, 7 days)
Default: 12 hours
This is the interval between wiper cleaning of the sensor window. It is recommended to start with
a setting of 30 minutes. This time may be adjusted according to the application. If readings
continue to be accurate, try a longer interval. If not, shorten the interval.

RESPONSE TIME

This is a damping function. While the Solitax takes readings continually, it will average them
together over the period of the response time. Once the response time has elapsed, the
displayed reading, 4-20 outputs, and alarm status are updated. (0 to 300 seconds)
Default: 3 seconds

LOGGER INTERVAL

This is the datalog interval, with options from 1–15 minutes. Values logged are the average of
the all readings during the previous logging interval. The controller will hold approximately
360 days of readings for one sensor at 15 minute intervals, or 24 days at 1 minute intervals (and
proportional in between). Default:10 minutes

SET DEFAULTS

Resets all user-editable options to the factory-defaults.

TEST/MAIN

PROBE INFO

Displays the sensor type, entered name of the sensor (Default: sensor serial number), the sensor
serial number, the software version number, and the sensor driver version number.

PROFILE

Select Profile Counter to display the number of wipes made (from 20000 backwards). Select
Reset Config to manually reset the profile counter.

COUNTER

Shows the number of hours or cycles left for operating hours, test/maint, gasket, and the motor.

TEST/MAIN

WIPE—Initiates the wiping action of the wiper.
SIGNALS—displays the signal outputs for the device.
OUTPUT MODE—Select the behavior of the instrument outputs (Hold, Active, Transfer,
Selection)
DEFAULT SETUP—Resets all user-editable options to the factory defaults.