Configure the sensor

Use the PROGNOSYS menu to monitor specific settings or to change the blink mode.

1. Push MAIN MENU>RTC MODULES/PROGNOSYS>PROGNOSYS.
2. Select the applicable sensor.
3. Select an option.

Option

Description

MEAS. INDICATOR

Shows the measurement indicator bar in %.

DETAILS

Shows the message list for the measurement indicator. Example: R<M—shows if
the reference signal is smaller than the measurement signal in %. MEAS EXT—
shows the absorbance value in %.

SERVICE INDICATOR Shows the number of days that remain until a maintenance task must be completed.

DETAILS

Shows the message list for the service indicator. Example: replace the wiper profile:
1 day or replace the seals: 42 days

DEVICE

Shows the sensor name.

LOCATION

Shows the location name where the sensor is used.

PROGNOSYS VERS.

Shows the PROGNOSYS software version.

BLINK MODE MEAS< Enters the measurement indicator value in %. The measurement indicator bar

flashes when the value decreases below the given value. Make sure to enter a
higher value when the measurement is related to a control system or a sensitive
parameter. Range: 0 to 100% (default = 0%)

BLINK MODE SERV<

Enters individual number of days for a maintenance task. The service indicator bar
flashes when the value decreases below the given value. Range: 0 to 200 days
(default = 0 days)

Service and measurement value indicator messages

W A R N I N G

For safe use of the instrument, obey the precautions and the instructions in the sensor documentation.

This list gives a general overview for the service and measurement value indicator messages. Refer
to the sensor documentation for a more detailed list of maintenance messages.

AMTAX sc and PHOSPHAX sc

Message

Possible cause

Solution

Instrument error

The instrument caused an error
message.

Examine the error message on the controller. Refer to
the sensor documentation for solutions. Make sure that
the error is shown in the service menu and push
START.

Instrument
warning

The instrument caused a warning
message.

Warm-up phase

The inside of the instrument is too
cold (e.g., the enclosure door was
open at low external
temperatures).

Wait until the warm-up phase is completed. If the start-
up temperature was higher, the warm-up phase can
vary from a few minutes to an hour.

Cooling down

The instrument is overheated and
is cooling-down. If the instrument
version has a filter probe, the
compressor is deactivated in this
phase.

• Wait until the cooling-down phase is completed.
• Make sure that no blockage of the vent is caused.
• Clean or change the air filter.
• Complete a function test on the fan.
• Specify the correct operating temperature.