Ldo probe, model 2 – Hach-Lange PROGNOSYS User Manual

Page 18

Advertising
background image

Message

Possible cause

Solution

Check wiper
function

The wiper is mechanically blocked. The
measurement window is not clean anymore.
No sample is pulled in the instrument. The
wiper is possibly in front of the window.

• Examine and clean the measurement

window fully.

• Complete a wiper test and do a stop

position test.

• Contact technical support.

Wiper position
unknown

Replace wiper
motor

The counter for wiper motor is expired.

Contact technical support to replace the
wiper motor.

Calibration data
faulty

The factory calibration data is lost.

Contact technical support.

Instrument error

The instrument caused an error message.

Examine the error message on the
controller. Refer to the sensor
documentation for solutions.

Instrument
warning

The instrument caused a warning message.

Humidity probe

Moisture in the sensor is over limit: the
desiccant is expired. There can be a problem
with the gasket set for the sensor sleeve or
the wiper axle.

Contact technical support immediately to
replace the gasket, so that the instrument
operates correctly.

LED faulty

LED intensity is too low.

Contact technical support.

Replace gasket

The time for the wiper axle gasket cycles is
expired.

Contact technical support to replace for a
wiper axle gasket.

LDO Probe, Model 2

Message

Possible cause

Solution

Red amplitude error

No sensor cap is installed.

Refer to the LDO manual to install the sensor
cap. If the cap is already installed, replace the
sensor.

The red LED is defective.

Contact technical support.

Blue amplitude error

No sensor cap is installed.

Refer to the LDO manual to install the sensor
cap. If the cap is already installed, replace the
sensor.

The blue LED is defective.

Contact technical support.

Temp <0°C/32°F

The sample temperature is less than
the specified value.

Make sure that the sample temperature is >
0°C (32°F). Move the sensor to another
position.

The thermistor system is defective.

Contact technical support.

Temp >50°C/122°F

The sample temperature is more than
the specified value.

Make sure that the sample temperature is <
50°C (122°F). Move the sensor to another
position.

The thermistor system is defective.

Contact technical support.