Stablcal standards storage, Turbidity measurement, Measurement notes – Hach-Lange 2100 AN IS Laboratory Turbidimeter Basic User Manual User Manual

Page 13

Advertising
background image

7. Remove the vial
from the sample cell
holder.

8. Do steps 5–10 for
the other StablCal vials
(from lowest to highest
NTU standard).
The mode display
shows "00" after the
last vial is measured.

9. Push CAL/Zero.
The instrument saves
the new calibration data
and goes back to
Measurement mode.

StablCal standards storage

• Do not move a StablCal standard to a different container for storage.

Keep StablCal standards in the plastic case provided with the cover
closed.

• Store at 5 to 25 °C (41 to 77 °F).
• For long-term storage (more than one month between use), keep at

5 °C (41 °F).

Turbidity measurement

W A R N I N G

Potential explosion and fire hazard. This instrument is for measuring water based
samples. Do not measure solvent or combustible based samples.

For accurate turbidity readings use clean sample cells and remove air
bubbles.

Measurement notes

Proper measurement techniques are important in minimizing the effects
of instrument variation, stray light and air bubbles. For accurate and
repeatable measurements:

Instrument

• Make sure that the instrument is on a level, stationary surface that is

free of vibration during the measurement.

• Instrument stabilization is immediate. No warm-up time is necessary.
• Always close the sample compartment lid during measurement,

calibration and storage.

• Remove the sample cell from the instrument and turn off the

instrument if the instrument is stored for an extended time period
(more than a month).

• Keep the sample compartment lid closed to keep dust and dirt out.

Sample cells

• Always cap the sample cell to prevent spillage of the sample into the

instrument.

• Always use clean sample cells in good condition. Dirty, scratched or

damaged cells can result in readings that are not accurate.

• Make sure that cold samples do not “fog” the sample cell.
• Store sample cells filled with distilled or deionized water and cap

tightly.

English 13