General information, Expanded manual version, Safety information – Hach-Lange 2100 AN IS Laboratory Turbidimeter Basic User Manual User Manual

Page 4

Advertising
background image

Specification

Details

Reading modes

Manual or auto range, signal averaging on and
adjustable or off, Ratio on or off

Power requirement

115–230 VAC, 50/60 Hz (automatic power selection)
28 W maximum

Pollution
degree/installation
category

2; II

Protection Class

1

Operating conditions

Temperature: 0 to 40 °C (32 to 104 °F)
Relative humidity: 0–90% at 25 °C, 0–75% at 40 °C,
noncondensing
Altitude: 2000 m (6560 ft) maximum
Indoor use only

Storage conditions

–40 to 60 °C (–40 to 140 °F), instrument only

Printer

Built-in (thermal, 58-mm, up to 28 column)

Interface

RS232C serial interface by way of DB9 subminiature D-
shell connector for data output to computer or printer,
and data input (command). No handshaking.

Air purge

Dry nitrogen or instrument grade air (ANSI MC 11.1,
1975)
0.1 scfm at 69 kPa (10 psig); 138 kPa (20 psig)
maximum
Hose barb connection for

1

/

8

-inch tubing

Sample cells

Round cells 95 x 25 mm (3.74 x 1 in.) borosilicate glass
with rubber-lined screw caps

Note: Smaller sample cells (less than 25 mm) can be used when
a cell adapter is used.

Sample requirements

25 mm sample cell: 20 mL minimum
0 to 95 °C (32 to 203 °F)

Enclosure

High-impact polycarbonate plastic

Dimensions

30.5 x 40 x 15.6 cm (12.0 x 15.7 x 6.1 in.)

Specification

Details

Weight

3.8 kg (8.5 lb)

Certification

CE, cETLus

1

Turbidity specifications identified using recently prepared formazin standard
and matched 25-mm sample cells.

2

Reference conditions: 23 ± 2 °C, 50% ± 10% RH noncondensing,
115/230 VAC, 50/60 Hz

3

Intermittent electromagnetic radiation of 3 volts/meter or greater may cause
slight accuracy shifts.

4

FNU is equivalent to NTU in the Ratio off mode.

General information

In no event will the manufacturer be liable for direct, indirect, special,
incidental or consequential damages resulting from any defect or
omission in this manual. The manufacturer reserves the right to make
changes in this manual and the products it describes at any time, without
notice or obligation. Revised editions are found on the manufacturer’s
website.

Expanded manual version

For additional information, refer to the CD for an expanded version of
this manual.

Safety information

N O T I C E

The manufacturer is not responsible for any damages due to misapplication or
misuse of this product including, without limitation, direct, incidental and
consequential damages, and disclaims such damages to the full extent permitted
under applicable law. The user is solely responsible to identify critical application
risks and install appropriate mechanisms to protect processes during a possible
equipment malfunction.

Please read this entire manual before unpacking, setting up or operating
this equipment. Pay attention to all danger and caution statements.
Failure to do so could result in serious injury to the operator or damage
to the equipment.

4 English

Advertising