Hach-Lange 2100 AN IS Laboratory Turbidimeter Basic User Manual User Manual

Page 3

Advertising
background image

Table of contents

Specifications

on page 3

Standard operation

on page 11

General information

on page 4

Maintenance

on page 16

User interface

on page 8

Troubleshooting

on page 17

Startup

on page 10

Specifications

Specifications are subject to change without notice.

Specification

Details

Measurement method

Nephelometric

Regulatory

Meets ISO 7027, DIN EN 27027, DIN 38404 and NFT
9033
ASTM D7315 - Standard Test Method for
Determination of Turbidity Above 1 Turbidity Unit (TU)
in Static Mode
ASTM D6655 - Standard Test Method for
Determination of Turbidity Below 5 NTU in Static Mode

Light source

Light-emitting diode (LED) at 860 ± 30 nm

Measurement modes

FNU, FAU, NTU, EBC, Abs (absorbance), %T (%
transmittance) and two user-defined units

Specification

Details

Range

FNU (manual range): 0–0.999, 0–9.99, 0–99.9, 0–1000
FNU (auto range): 0–1000
FAU (manual range): 20–99.9, 20–10,000
FAU (auto range): 20–10,000
NTU (Ratio on, manual range): 0–0.999, 0–9.99, 0–
99.9, 0–10,000
NTU (Ratio on, auto range): 0–10,000 auto decimal
NTU (Ratio off): 0–40
EBC (Ratio on, manual range): 0–0.999, 0–9.99, 0–
99.9, 0–2450
EBC (Ratio on, auto range): 0–2450 auto decimal
EBC (Ratio off): 0–9.8
Absorbance (manual range): 0–0.999, 0–2.00
Absorbance (auto range): 0–2.00
Transmittance (%): 1.0–100

Accuracy

1,

2,

3

FNU

4

: ±2% of reading plus 0.01 FNU from 0–1000 FNU

FAU: ±10% of reading from 20–10,000 NTU
NTU

4

: ±2% of reading plus 0.01 NTU from 0–

1000 NTU, ±5% of reading from 1000–4000 NTU,
±10% of reading from 4000–10,000 NTU
Absorbance: ±0.005 Abs from 0–1 Abs at 860 nm
Transmittance: 0.12% T from 10–100% T at 860 nm

Resolution

Turbidity: 0.001 FNU/NTU/EBC
Absorbance: 0.001 Abs
Transmittance: 0.1% T

Repeatability

±1% of reading or 0.01 FNU, whichever is greater
(under reference conditions)

Response time

Signal averaging off: 6.8 seconds
Signal averaging on: 14 seconds (when
10 measurements are used to calculate the average)

Stabilization time

Immediately