Configure the rs232 connection, Computer (rs232) commands, Connect to a data recorder – Hach-Lange 2100 AN IS Laboratory Turbidimeter User Manual User Manual

Page 38

Advertising
background image

5. Push ENTER.
6. Push SETUP.

Configure the RS232 connection

1. Push SETUP. The SETUP light turns on.
2. Use the arrow keys to select an option:

Option

Description

10

Sets the baud rate (default=1200).

11

Sets the character length (default=8).

12

Sets the stop bit (default=1).

13

Sets the parity select (default=NONE).

3. Push ENTER.
4. Use the arrow keys to change the value.
5. Push ENTER.
6. Push SETUP.

Computer (RS232) commands

A communication program (i.e., such as Window Terminal or ProComm
Plus) is recommended for computer operation of the instrument.
Configure the communication program to the RS232 connection settings.
Refer to

Configure the RS232 connection

on page 36.

Table 5

shows the RS232 command set for the instrument.

Table 5 RS232 command set

Command Description

VAL

Gets the current measurement with the measurement units.

LST

Gets the calibration standards and coefficients.

DAT

Gets the current date.
To change the date, enter DAT=MM/DD/YY (MM=month, DD=day,
YY=year), then push Enter.

Table 5 RS232 command set (continued)

Command Description

TIM

Gets the current time in 24-hour format.
To change the time, enter TIM=HH:MM (HH=hour, MM=minutes),
then push Enter.

RMN

Gets the recorder minimum value.
To change the recorder minimum value, enter RMN=XXXXX
(XXXXX=a number, minimum value=0), then push Enter.

RMX

Gets the recorder maximum value.
To change the recorder maximum value, enter RMX=XXXXX
(XXXXX=a number, maximum value=10,000), then push Enter.

RTN

Gets the recorder trim minimum value.
To change the recorder minimum value, enter RTN=XXXXX
(XXXXX=a number, minimum value=200), then push Enter.

RTX

Gets the recorder trim maximum value.
To change the recorder maximum value, enter RTX=XXXX
(XXXX=a number, maximum value=4800), then push Enter.

SAV

Gets the signal average buffer size.
To change the signal average buffer size, enter SAV=XX (XX=a
number, maximum value=15, default=10), then push Enter.

Connect to a data recorder

Note: Use a twisted-pair, shielded recorder cable. Use of non-shielded recorder
cable may result in radio wave emission levels greater than is allowed under the
compliance regulations listed.
Note: Connect the shield of the recorder cable to the recording device chassis
ground terminal to decrease the effects of unwanted interferences.

Connect a ¼–inch recorder phone plug to the recorder output jack on the
back of the instrument. Refer to

Figure 2

on page 8. For the best

performance, use a twisted-pair, shielded recorder cable that is no more
than 1.8 m (6 ft) in length with a load impedance greater than 10 kohms.

36 English